1) Instrument displays "TANK FLOW LOW".
   AMCC is not detecting an appropriate flow of gas from the standard. Check to see if the gas tank is connected and if it has gas in it.
AMCC

2) Instrument Displays "SUBJECT TESTS DISABLED".
   "Either a routine accuracy check was not performed or the accuracy check failed. If the check did not occur, an accuracy check must be performed before further subject testing. If the check failed, a calibration followed by a successful accuracy check must be performed."