10) Depress the SET button. Your calibration is complete.
  • SET button depressed
  • Observe final reading
  • Target value of the standard used for this demonstration was .080%, the instrument is calibrated

 

 

11)  It is essential to verify the calibration. Wait 3 minutes then run an Accuracy check using a new mouthpiece. (Follow the step by step procedures described in the Post Calibration Accuracy Check).

The Accuracy Check reading following a Calibration should be within + .003 of the target value of the Calibration Standard. If it is not, wait another three minutes and then repeat the Calibration Procedure followed by an Accuracy Check.

 

Post Calibration Accuracy Check

Back          Procedure Overview