10) Depress the SET
button. Your calibration is complete.
- SET button depressed
- Observe final reading
- Target value of the standard used for this
demonstration was .080%, the instrument is calibrated
|
|
|
11) It is essential to verify the calibration. Wait 3 minutes then run an Accuracy check using a new mouthpiece. (Follow the step by step procedures described
in the Post Calibration Accuracy Check).
The Accuracy Check reading following a Calibration should be within + .003 of the target value of the Calibration Standard. If it is not, wait another three minutes and then repeat the Calibration Procedure followed by an Accuracy Check.
|
|
| Post Calibration Accuracy Check
|
| Back Procedure Overview |
|
|