ASIV Calibration
Prior to calibrating an Intoximeters breath alcohol testing instrument,
you should be trained to perform the calibration functions as allowed under
your specific testing program. The calibration procedures below
are designed to help you ensure precise and accurate test results. If
according to your program you are not authorized to perform calibrations,
please see your program administrator or contact Intoximeters for
assistance. Typically, a calibration is not required if the results of
accuracy checks are within the programs acceptable tolerance.
The procedure for calibrating an Intoximeters breath alcohol testing
device will vary based on instrument model and type. Refer to the
Operator's Manual for your instrument for specific instructions on
calibrating your device. DO NOT perform this procedure unless
you are authorized to do so.
Getting Started
Below are basic calibration instructions for the Intoximeters Alco Sensor
IV breath testing instrument. Unless your testing procedure dictates
otherwise, a calibration is only necessary if a accuracy check has proven
the instrument to be out of tolerance. Some users choose to calibrate
the instrument more frequently to ensure that the instrument remains within
tighter tolerances than required by their quality assurance plan and this
process will not damage the instrument.
Tools Required: Before beginning
have items available: instrument, calibration standard, new mouthpiece,
calibration tool
|