An accuracy check should be performed periodically. The
frequency should be determined by your Quality Assurance Program.
Prior to running an Accuracy Check you should make certain
that the Standard you will be using is current and all of the operating
parameters for that device have been met:
- Have as little tubing as possible between the gas out
port of the standard and the mouthpiece of the Alco Sensor IV.
- Check Expiration Date (on Wet Bath Simulator Solution or
Dry Gas Tank)
- Check Temperature of the standard (Check Wet Bath
Simulator Temperature)
- Purge the simulator or dry gas regulator for several
seconds
- Know the current expected value of the
Standard gas
- If you keep a calibration log, it should be present
Tools Required: Before beginning
have items available: instrument, calibration standard, new mouthpiece,
calibration tool
|