Accuracy Check - ASIII
1) Check temperature strip on back of Alco-Sensor III –
any visible number verifies that the instrument is at proper operating
temperature.
|
Temperature window display
|
|
2) Depress read button, .000 should be seen for at least 5 to 10
seconds.
|
READ button depressed
ASIII displays .000
|
|
3) If .000 is not seen, depress SET button and recheck in one minute.
(Follow Step 2 again. The unit must display .000 for 5 to 10 seconds before
proceeding with the Accuracy Check).
4) Depress set button on Alco-Sensor III.
|
 SET
Button Pressed
|
|
5) Follow instructions on the dry gas standard to mount the regulator.
6) If this is the first test from the dry gas standard today, purge
regulator by depressing button/valve on regulator for approximately 3 to 4
seconds.
7) Attach a mouthpiece to the Alco-Sensor III and then to the
Standard's output port.
|
|
|