6) Depress the regulator control button.
7) When there is a continuous airflow - 6 liters per minute
for four seconds - the instrument should capture a sample automatically.
Make certain that the gas continues to flow for 1-2 seconds after the
sample is captured. (Sample collection is identified by a clicking sound.)
(Note: If for some reason you
cannot provide an adequate flow rate for the instrument to collect an
automatic sample, it is possible to perform an Accuracy Check by taking a
manual sample. To perform an Accuracy Check in this manner, depress
the regulator control button and have the gas flowing to the instrument for
seven seconds. On the 5th second depress the ON button to take
a manual sample. Make certain that the gas continues to flow for 1-2
seconds after the sample is taken.) |