Alco-Sensor FST Accuracy Check
An Accuracy Check should be performed periodically. The frequency should
be determined by your Quality Assurance Program.
Prior to running an Accuracy Check you should make certain that the
Standard you will be using is current and all of the operating parameters
for that device have been met:
- Have as little tubing as possible between the gas out port of the
standard and the mouthpiece of the Alco-Sensor FST.
- If using a Wet Bath or Dry Gas Standard check the expiration date
- If using a Wet Bath Standard check the temperature
- Purge the simulator or dry gas regulator for several seconds
- Know the current expected value of the Standard gas by using
a True-Cal or Elevation Chart
- If you keep a calibration log, one should be maintained for each FST
Tools Required: Before beginning
have items available: instrument, calibration standard, new mouthpiece
|