Sensor
The fuel
cell sensor generates a response that is proportional to the
Breath Alcohol Concentration.
Specificity
The fuel
cell sensor is sensitive to alcohol. It does not respond to
acetone or other substances which are found in the breath.
Measurement Limits
The Alco-Sensor can accurately detect
breath alcohol levels between .00 - .40 BrAC.
Response Time
The unit responds within 10 seconds on
negative samples and within 30 - 45 seconds on positive samples.
Sampling Rate
The Alco-Sensor can run negative tests every 15 to 20 seconds.
Positive tests can be run every two minutes.
Drift
Instrument will maintain calibration
(plus or minus .01 at the .10 level) for months.
Calibration
Checks and calibrations should be
performed with either a National Highway Traffic Safety
Administration's (NHTSA) approved wet bath simulator or a dry gas
standard.
Environmental
Can operate between 0° and 40° C.